Measurement Accuracy |
Voltage DC |
30.000 V |
0.01% + 2 mV (upper display) |
20.000 V |
0.01% + 2 mV (lower display) |
90.00 mV |
0.01% + 10 μV |
-10.00 mV to 75.00 mV |
0.025% + 1 count (via TC connector) |
Current DC |
24.000 mA |
0.01%+ 2 mA |
Resistance |
0.0 to 4000.0 Ω |
± 0.015% |
Frequency |
2.0 to 1000.0 CPM |
0.05% + 0.1 cpm |
1.0 to 1000.0 Hz |
0.05% + 0.1 Hz |
1.00 to 15.00 kHz |
0.05% + 0.01 kHz |
Sensitivity |
1 V peak-to-peak-minimum |
Pressure |
Accuracy from 0.025% of range using any of 29 pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure.
|
Source Accuracy |
Voltage DC |
100.00 mV |
0.01% + 10 μV |
20.000 V |
0.01% + 2 mV |
-10.00 mV to 75.00 mV |
0.025% + 1 count (via TC connector) |
Current DC |
24.000 mA (Source) |
0.01% + 2 μA |
24.000 mA (Simulate) |
0.01% + 2 μA |
Resistance |
Range, accuracy (percent of reading + floor) and excitation current |
5 to 400 Ω, 0.015% + 0.1 Ω, 0.1 to 0.5 mA |
5 to 400 Ω, 0.015% + 0.05 Ω, 0.5 to 3 mA |
400 to 1500 Ω, 0.015% + 0.5 Ω, 0.05 to 0.8 mA |
1500 to 4000 Ω, 0.015% + 0.5 Ω, 0.05 to 0.4 mA |
Frequency |
2.0 to 1000.0 CPM |
0.05% |
1.0 to 1000.0 Hz |
0.05% |
1.0 to 10.0 kHz |
0.25% |
10.0 to 15.0 kHz |
0.5% |
Waveform |
5 V p-p squarewave, -0.1 V offset |
RTDs and Thermocouples |
Measure and source accuracy |
Cu10 |
1.8°C |
NI-120 |
0.15°C |
PT-100 (385) |
0.45°C |
PT-100 (392) |
0.35°C |
PT-100 (391-JIS) |
0.35°C |
PT-200 (385) |
0.95°C |
PT-500 (385) |
0.55°C |
PT-1000 (385) |
0.35°C |
RTD Resolution |
0.01°C (0.01°F) |
Cu10.01 C |
0.1°C (0.1°F) |
J |
0.4°C |
K |
0.5°C |
T |
0.4°C |
E |
0.4°C |
R |
1.4°C |
S |
1.4°C |
B |
1.4°C |
L |
0.45°C |
U |
0.45°C |
N |
0.6°C |
XK |
0.4°C |
BP |
1.1°C |
TC Resolution |
J, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F |
Specifications |
Ramp functions |
Source functions |
Voltage, current, resistance, frequency, temperature |
Ramps |
Slow ramp, Fast ramp, 25% step-ramp |
Loop power function |
Voltage |
24 V |
Accuracy |
10% |
Maximum current |
22 mA, short circuit protected |
Step functions |
Source functions |
Voltage, current, resistance, frequency, temperature |
Steps |
25% of range, 100% of range |
Environmental Specifications |
Operating temperature |
-10°C to +55°C |
Storage temperature |
-20°C to +71°C |
Operating altitude |
3000 m |
Safety Specifications |
Agency approvals |
EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992 |
Mechanical and General Specifications |
Size |
200 x 96 x 47 mm (7.9 x 3.8 x 1.9 in) |
Weight |
650 g (23 oz) |
Batteries |
4 AA alkaline batteries |
Battery replacement |
Separate battery compartment, accessible without breaking calibration seal |
Side port connections |
Pressure module connector, also used for remote real-time programming |