Fluke 7526A Precision Process Calibrator
Supplier: Iconic Engineering Limited
Brand: Fluke
Category: Calibration
Sub Category: Multifunction Calibrators
Out of Stock Ask for PriceModel: Fluke 7526A Precision Process Calibrator
Fluke 7526A Precision Process Calibrator “Doing more with less” is a requirement every process manufacturer faces today. To maintain product quality, reduce waste, improve efficiency and conform to regulatory standards, instruments that measure process variables such as temperature and pressure must be calibrated at regular intervals.
Selecting the right calibrator for the job, however, can be tricky—especially when attempting to balance precision and versatility with the cost. Less expensive than high-end multi-product calibrators, yet more precise and versatile than handheld field calibrators, the 7526A is a “best fit” for instrument shops that value precision, versatility, and economy.
Intuitive user interface
Fluke 7526A The user interface includes cursor controls, function keys, and a ten-key keypad, making it easy to navigate through intuitive menus; store and recall up to nine setpoints for each input/output parameter; enter RTD or SPRT coefficients, and easily change display units with a keypress. With two LCD displays, you can easily view both source and measurement results simultaneously.
However, Iconic engineering suggest you as an authorized distributor of Fluke corporation Permanently you can use Fluke product and when you need an order from Tools Bangladesh.
Features
Key Features: Fluke 7526A Precision Process Calibrator
- Sources and measures dc voltage, current, resistance, RTDs and thermocouples
- Extend pressure measurement range with any Fluke 750P and 700P Series Pressure Modules
- Includes 24 V dc transmitter loop power supply
- Measures 4-20 mA loop current
- Includes automated switch-test function
- Accepts ITS-90 coefficients for accurate SPRT measurements
- Compatible with MET/CAL® Calibration Software
- Includes certificate of calibration traceable to national standards (optional A2LA accredited calibration available upon request)
Specifications
Specifications: Fluke 7526A Precision Process Calibrator
DC voltage, output | |||
Range [1] | Absolute Uncertainty, ± (ppm of output + μV), 1-yr |
Resolution | |
0 to 100 mV | 30 | 3 | 1 μV |
0 to 1 V | 30 | 10 | 10 μV |
0 to 10 V | 30 | 100 | 100 μV |
0 to 100 V | 30 | 1 mV | 1 mV |
TC Output and Input | |||
–10 mV to 75 mV | 30 | 2 μV | 1 μV |
[1] All outputs are positive only unless otherwise noted
DC voltage, isolated input | |||
Range | Absolute Uncertainty, ± (ppm of reading + mV), 1-yr |
Resolution | |
0 to 10 V | 100 | 0.2 | 100 μV |
10 V to 100 V | 100 | 2.0 | 1 mV |
DC Current, output | |||
Range[1] | Absolute Uncertainty, ± (ppm of reading + μA), 1-yr |
Resolution | |
0 to 100 mA | 50 | 1 | 1 μA |
[1] For line voltages less than 95 V, ±100 ppm of reading
DC current, isolated input | |||
Range | Absolute Uncertainty, ± (ppm of reading + μA), 1-yr |
Resolution | |
0 mA to 50 mA | 100 | 1 | 0.1 μA |
0 mA to 24 mA [1][2] (Loop Power) |
100 | 1 | 0.1 μA |
[1] Loop Power: 24 V ±10%
[2] HART Resistor: 250 Ω ± 3%
Resistance, output | |||
Range | Absolute Uncertainty, tcal ± 5 °C, ± Ohms, 1-yr |
Resolution | Nominal current |
5 Ω to 400 Ω | 0.015 | 0.001 Ω | 1 to 3 mA |
5 Ω to 4 kΩ | 0.3 | 0.01 Ω | 100 μA to 1 mA |
Resistance, input | |||
Range | Absolute Uncertainty, ± (ppm of reading + Ω), 1-yr |
Resolution | |
0 Ω to 400 Ω | 20 | 0.004 | 0.001 Ω |
0 Ω to 4 kΩ | 20 | 0.04 | 0.01 Ω |
Sample thermocouple accuracy, input/output (does not include all available TC types)[1] |
|||
TC type | Temperature Range (°C) | Absolute Uncertainty, tcal ±5 °C, ± (°C), 1-yr [2] |
|
Min | Max | ||
J | -210 | 1200 | 0.09 |
K | -250 | 1372 | 0.1 |
S | -50 | 1767 | 0.29 |
T | -250 | 400 | 0.11 |
[1] See extended specifications for all TC types (B,C,E,J,K,L,N,R,S,T,U,XK,BP)
[2] Best accuracy within specified TC temperature range
Sample RTD and thermistor, output (does not include all available RTD types)[1] |
|||
RTD Type | Temperature Range (°C) | Absolute Uncertainty, tcal ±5 °C, ± (°C), 1-yr |
|
Min | Max | ||
Pt 385, 100 Ω | -200 | 800 | 0.05 |
YSI 400 | 15 | 50 | 0.007 |
[1] See extended specifications for all RTD types: Pt-100 (385, 3926, 3916), Pt- 200, Pt-500, Pt-1000, Ni-120, Cu-427, SPRT
Sample RTD and thermistor, intput (does not include all available RTD types)[1] |
|||
RTD Type | Temperature Range (°C) | Absolute Uncertainty, tcal ±5 °C, ± (°C), 1-yr |
|
Min | Max | ||
Pt 385, 100 Ω | –80 | 100 | 0.020 |
100 | 300 | 0.024 | |
300 | 800 | 0.038 | |
YSI 400 | 15 | 50 | 0.007 |
[1] See extended specifications for all RTD types: Pt-100 (385, 3926, 3916), Pt- 200, Pt-500, Pt-1000, Ni-120, Cu-427, SPRT
General specifications | ||
Standard interface | RS-232, IEEE-488 (GPIB) | |
Temperature performance |
Operating: | 0 °C to 50 °C |
Calibration (tcal): | 18 °C to 28 °C | |
Storage: | –20 °C to 70 °C | |
Electromagnetic compatibility |
CE: Conforms to EN61326; operation in controlled EM environments |
|
Temperature coefficient | Temperature coefficient for temperatures outside tcal 5 °C is 10% of the 90-day specification (or 1 year if applicable) per °C |
|
Relative humidity | Operating: | <80% to 30 °C |
<70% to 40 °C | ||
<40% to 50 °C | ||
Altitude | Operating: | 3,000 m (9,800 ft) max |
Non-operating: | 12,200 m (40,000 ft) max | |
Safety | EN/IEC
61010-1:2010 3
rd Edition,UL 61010-1:2012, CAN/CSA 22.2 No. 61010-1-12 |
|
Analog low isolation | 20 V | |
Line power | 120 V~: | 100 V to 120 V |
240 V~: | 220 V to 240 V | |
Line frequency | 47 Hz to 63 Hz | |
Line voltage variation | ± 10 % about setting | |
Power consumption | 15 VA maximum | |
Dimensions | Height: | 14.6 cm (5.75 in) |
Width: | 44.5 cm (17.5 in) | |
Depth: | 29.8 cm (11.75 in) | |
Weight (without options) |
4.24 kg (9.35 lb) |
Includes
Includes: Fluke 7526A Precision Process Calibrator
Model Name | Description |
---|---|
7526A |
Precision Process Calibrator Includes traceable calibration report, user manual CD, getting started guide, power cord, thermocouple shorting jumper and USB-to-serial adapter cable. |
Accessories
Accessories: Fluke 7526A Precision Process Calibrator & common to all models
Accessory | Description |
---|---|
Y7526A |
Rack Mount Kit
|
7526A-CASE |
Carrying Case for 7526A |
700LTP-1 |
Low-pressure Test Pump |
700HTH-1 |
Fluke 700HTH-1 Hydraulic Test Hose |
700HTP-2 |
Hydraulic Test Pump |
71X |
Hose Kit Accessory |
700ILF |
In-line Filter |
700TC1 |
Thermocouple Plug Kit 700C1 |
700TC2 |
Thermocouple Plug Kit 700TC2 |
Fluke 750P Series Pressure Modules |
Pressure Module Series |
5520A-525A/LEADS |
Test Lead Set |
Resources

5502A Multi-Product..
Model: 5502A Multi-Product Calibrator F..

5522A Multi-Product..
The 5522A Multi-Product Calibrator repla..

732C and 734C DC Vol..
Model: 732C and 734C DC Voltage Referenc..

5730A High Performan..
Model: 5730A High-Performance Multifunct..

5790B AC Measurement..
The 5790B covers an alternating voltage..

5322A Electrical Saf..
The 5322A combines many functions into a..

910/910R GPS Frequen..
The 910 and 910R GPS frequency reference..

Fluke 710 mA Loop Va..
Model: Fluke 710 mA Loop Valve Tester F..