Model: Fluke 725 Multifunction Process Calibrator
Fluke 725 Multifunction Process Calibrator you can easily source and measure almost any process parameter and test almost any process device. Measure and source mA, volts, temperature (RTDs and thermocouples), frequency, ohms, and pressure, using optional pressure modules.
From the moment you pick it up, the Fluke 725 is ready to perform as a powerful, yet easy-to-use field calibrator. Its simple controls, without menus, make operation easy. If you want to get it you can order not at Tools Bangladesh.
Key features: Fluke 725 Multifunction Process Calibrator
- Powerful, easy-to-use field calibrator
- Measures volts, mA, pressure, RTDs, thermocouples, frequency, and ohms to test sensors and transmitters
- Sources/simulates volts dc, mA, thermocouples , RTDs, frequency, ohms, and pressure to calibrate transmitters
- Measures/sources pressure using any of 50 Fluke 750P Pressure Modules
- Easy to read measure/source screen displays input and output simultaneously
Specifications: Fluke 725 Multifunction Process Calibrator
| Measurement Accuracy |
| Voltage DC |
30.000 V |
0.02% + 2 counts (upper display) |
| 30.000 V |
0.02% + 2 counts (lower display) |
| 100.00 mV |
0.02% + 2 counts |
| -10.00 mV to 75.00 mV |
0.025% + 1 count (via TC connector) |
| Current DC |
24.000 mA |
0.02% + 2 counts |
| Resistance |
0.0 to 400.0 Ω |
0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire) |
| 401 to 1500 Ω |
0.5 Ω (4-wire), 1 Ω (2- and 3-wire) |
| 1500 to 3200 Ω |
1 Ω (4-wire), 1.5 Ω (2- and 3-wire) |
| Frequency |
2.0 to 1000.0 CPM |
0.05% + 1 count |
| 1.0 to 1100.0 Hz |
0.05% + 1 count |
| 1.00 to 10.00 kHz |
0.05% + 1 count |
| Sensitivity |
1 V peak-to-peak-minimum |
| Pressure |
Accuracy from 0.025% of range using any of 29 pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure. |
| Source Accuracy |
| Voltage DC |
100.00 mV |
0.02% + 2 counts |
| 10.000 V |
0.02% + 2 counts |
| -10.00 mV to 75.00 mV |
0.025% + 1 count (via TC connector) |
| Current DC |
24.000 mA (source) |
0.02% + 2 counts |
| 24.000 mA (simulate) |
0.02% + 2 counts |
| Resistance |
15.0 to 400.0 Ω |
0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA) |
| 401 to 1500 Ω |
0.5 Ω (excitation current 0.05 to 0.8 mA) |
| 1500 to 3200 Ω |
1 Ω (excitation current 0.05 to 0.4 mA) |
| Frequency |
2.0 to 1000.0 CPM |
0.05% |
| 1.0 to 1100.0 Hz |
0.05% |
| 1.00 to 10.00 kHz |
0.25% |
| Waveform |
5 V p-p squarewave, -0.1 V offset |
| RTDs and Thermocouples |
| Measure accuracy |
NI-120 |
0.2°C |
| PT-100 (385) |
0.33°C |
| PT-100 (393) |
0.3°C |
| PT-100 (JIS) |
0.3°C |
| PT-200 (385) |
0.2°C |
| PT-500 (385) |
0.3°C |
| PT-1000 (385) |
0.2°C |
| Resolution |
0.1°C |
| J |
0.7°C |
| K |
0.8°C |
| T |
0.8°C |
| E |
0.7°C |
| R |
1.8°C |
| S |
1.5°C |
| B |
1.4°C |
| L |
0.7°C |
| U |
0.75°C |
| N |
0.9°C |
| Resolution |
J, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F |
| XK |
0.6°C |
| BP |
1.2°C |
| Source accuracy |
NI-120 |
0.2°C |
| PT-100 (385) |
0.33°C |
| PT-100 (393) |
0.3°C |
| PT-100 (JIS) |
0.3°C |
| PT-200 (385) |
0.2°C |
| PT-500 (385) |
0.3°C |
| PT-1000 (385) |
0.2°C |
| Resolution |
0.1°C |
| Accuracy stated for 4-wire measurement. |
| J |
0.7°C |
| K |
0.8°C |
| T |
0.8°C |
| E |
0.7°C |
| R |
1.4°C |
| S |
1.5°C |
| B |
1.4°C |
| L |
0.7°C |
| U |
0.75°C |
| N |
0.9°C |
| Resolution |
J, K, T, E, L, N, U: 0.1°C, B, R, S: 1°C |
| XK |
0.6°C |
| BP |
1.2°C |
| Specifications |
| Ramp functions |
Source functions |
Voltage, current, resistance, frequency, temperature |
| Ramps |
Slow ramp, Fast ramp, 25% step-ramp |
| Loop power function |
Voltage |
24 V |
| Accuracy |
10% |
| Maximum current |
22 mA, short circuit protected |
| Step functions |
Source functions |
Voltage, current, resistance, frequency, temperature |
| Steps |
25% of range, 100% of range |
| Environmental Specifications |
| Operating temperature |
-10°C to +55°C |
| Storage temperature |
-20°C to 71°C |
| Operating altitude |
3000 m |
| Safety Specifications |
| Agency approvals |
EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992 |
| Mechanical and General Specifications |
| Size |
130 x 236 x 61 mm |
| Weight |
0.65 kg |
| Batteries |
4 AA alkaline batteries |
| Battery replacement |
Separate battery compartment, accessable without breaking calibration seal |
| Side port connections |
Pressure module connector, also used for remote real-time programming |
Includes: Fluke 725 Multifunction Process Calibrator
- One set of AC72 alligator clips
- One set of stackable test leads
- One set of TL 75 testleads
- User manuals
- NIST- traceable calibration report and data
Accessories: Fluke 725 Multifunction Process Calibrator
Test leads (1)
- Fluke TLK-220 US SureGrip™ Industrial Test Lead
Cases (3)
- Fluke C100 Universal Carrying Case
- Fluke C120 Hard Carrying Case
- Fluke C125 Meter Case
Clips (1)
- Fluke AC220 SureGrip™ Alligator Clips
Lights (1)
- Fluke L215 SureGrip™ Kit with Probe Light and Probe Extenders
Pressure (1)
- Fluke 700PRV Pressure Relief Valve Kit
Pumps (1)
- Fluke 700 PMP Pressure Pump
Temperature (1)
- Fluke 700TC2 Thermocouple Plug Kit 700TC2