Kyoritsu 2433R is a current leakage clamp meter, taking into account the triumph to produce a result AC current taking place to 400A across three scales (4/40mA/400A). With genuine RMS The tester features a frequency selector switch to eliminate the effect of harmonics. This provides more accurate measurement, especially at the low fade away from the current scale. The 2433R model as well as has an elevation child support takes motion, to enable summit value measurements.
Featuring a 0.01mA loud, the 2433R clamp tester provides accurate current readings. A rotary switch in a bank account to the order of the device allows easy-one finger gift-upon and range selection, allowing the adherent to test efficiently. Standard Kyoritsu features such as the data-bond progression and auto-knack off are included.
Key Features: Kyoritsu 2433R Clamp Meters
• Frequency Selector Switch to eliminate the effect of harmonics.
• Three AC current ranges:40mA/400mA/400A.
• Resolution 0.01mA.
Specifications: Kyoritsu 2433R Clamp Meters
| AC A (50/60Hz) |
40/400mA/400A
±1%rdg±5dgt (40/400mA)
±1%rdg±5dgt (0 - 300A)
±2%rdg (300.1 - 399.9A) |
| AC A (WIDE) |
40/400mA/400A
±1%rdg±5dgt [50/60Hz],
±2.5%rdg±10dgt [20Hz - 1kHz] (40/400mA)
±1%rdg±5dgt [50/60Hz],
±2.5%rdg±10dgt [40Hz - 1kHz] (0 - 300A)
±2%rdg [50/60Hz],
±5%rdg [40Hz - 1kHz] (300.1 - 399.9A) |
| Maximum circuit voltage |
600V AC/DC (between line/neutral)
300V AC/DC (against earth) |
| Conductor size |
Ø40mm max. |
| Frequency response |
20Hz - 1kHz [40Hz - 1kHz:400A] |
| Effect of External Stray Magnetic Field |
10mA AC approx. in proximity to a 15mm-dia
conductor carrying 100A AC |
| Applicable standards |
IEC 61010-1 CAT III 300V Pollution Degree 2
IEC 61010-2-032 |
| Response Time |
Approx. 2 seconds |
| Power source |
Two R03 or equivalent (DC1.5V) batteries |
| Dimensions |
185(L) × 81(W) × 32(D)mm |
| Weight |
270g approx. |
| Included Accessories |
9052 (Carrying Case) R03 (1.5V) × 2
Instruction Manual |