| DC Voltage range | 
			60.000 mV to 1000.0 V, 6 ranges, Basic accuracy: ±0.025 % rdg. ±2 dgt. | 
		
		
			| AC Voltage* range | 
			60.000 mV to 1000.0 V, 6 ranges, Frequency characteristics: 20 Hz - 100 kHz 
			Basic accuracy 45 - 65 Hz : ±0.2 % rdg. ±25 dgt. (True RMS, crest factor 3) | 
		
		
			| DC + AC Voltage* range | 
			6.0000 V to 1000.0 V, 4 ranges, Frequency characteristics: 20 Hz - 100 kHz 
			Basic accuracy 45 - 65 Hz : ±0.3 % rdg. ±30 dgt. (True RMS, crest factor 3) | 
		
		
			| Resistance range | 
			60.000 Ω to 600.0 MΩ, 8 ranges, Basic accuracy: ±0.03 % rdg. ±2 dgt. | 
		
		
			| DC Current range | 
			600.00 μA to 600.00 mA, 4 ranges 
			Basic accuracy: ±0.05 % rdg. ±5 dgt. | 
		
		
			| AC Current* range | 
			600.00 μA to 600.00 mA, 4 ranges 
			Basic accuracy 45 - 65 Hz : ±0.6 % rdg. ±5 dgt. (True RMS, crest factor 3) 
			Frequency characteristics: 20 Hz - 20 kHz (at 600 μA to 600 mA range) | 
		
		
			AC Current* range 
			(use with Clamp on probes) | 
			10.00 A to 1000 A, 7 ranges 
			Add the Clamp on probe accuracy to Basic accuracy 40 - 65 Hz : ±0.6 % rdg. ±2 dgt. (True RMS, crest factor 3) | 
		
		
			| Peak | 
			DC V measurement: Signal width 4 msec or more (single), 1 msec or more (repeated) 
			AC V, DC/AC A measurement: 
			Signal width 1 msec or more (single), 250 μsec or more (repeated) | 
		
		
			| Capacitance range | 
			1.000 nF to 100.0 mF, 9 ranges, Basic accuracy: ±1.0 % rdg. ±5 dgt. | 
		
		
			| Continuity Check | 
			Continuity threshold: 20/50/100/500 Ω, Response time: 10 ms or more | 
		
		
			| Diode test | 
			Open terminal voltage: 4.5 V or less, Testing current 1.2 mA or less, Threshold of forward voltage: 0.15 V to 3 V, seven stages | 
		
		
			| Frequency range | 
			AC V, DC+AC V, AC A measurement, at pulse width 1 μs or more (50 % duty ratio) 
			99.999 Hz (0.5 Hz or more) to 500.00 kHz, 5 ranges, ±0.005 % rdg. ±3 dgt. | 
		
		
			| dB conversion | 
			Standard impedance setting (dBm), 4 Ω to 1200 Ω, 20 stages 
			Display dB conversion value of AC voltage (dBV) | 
		
		
			| Temperature (thermocouples) | 
			K: -40.0 °C to 800.0 °C (-40.0 °F to 1472.0 °F) 
			Add accuracy of the Thermocouple probe to main unit accuracy: ±0.5 % rdg. ±3 °C | 
		
		
			| Other functions | 
			Filter function (remove harmonic noise, use only at 600 V AC, 1000 V AC ranges), display value hold, auto hold, MAX/MIN value display, PEAK value display, relative value display, sampling setting, internal memory (400 data), auto-power save, USB communication (option), mis-insertion prevention shutters, decibel conversion, 4-20 mA percentage conversion | 
		
		
			| Display | 
			Main and sub displays: 5-digits LCD, max. 60000 digits | 
		
		
			| Display refresh rates | 
			5 times/s (Capacitance measurement: 0.05 to 2 times/s, depending on measured value, Temperature: 1 time/s ) | 
		
		
			| Power supply | 
			LR6 (AA) alkaline batteries × 4, Continuous use: 100 hours | 
		
		
			| Dimensions and mass | 
			93 mm (3.66 in) W × 197 mm (7.76 in) H× 53 mm (2.09 in) D, 650 g (22.9 oz) (with test leads holder and batteries) | 
		
		
			| Included accessories | 
			Test Lead L9207-10 × 1, Instruction Manual × 1, LR6 Alkaline Battery × 4 |