To ensure the safe use of electrical products, electrical safety tests such as insulation resistance, withstand voltage, ground resistance, and leakage current must always be conducted. Hioki Leakage Current Testers comply with laws and standards regarding medical electrical equipment as well as non-medical electrical equipment, and can be used to measure leakage current in all types of electrical products from computers to medical equipment. The ST5540 consists of measurement networks that simulate the human body and a high-frequency volt meter, and is also capable of switching power supply polarities and making measurements under simulated single-fault conditions in equipment under test. HIOKI LEAK CURRENT HiTESTER ST5540 Price in Bangladesh is competitive for precision insulation and leakage current testing. Ideal for industrial use, it ensures high accuracy and safety. Explore top features and reliable performance tailored for the Bangladeshi market. Get the best value with trusted quality and efficiency for your applications.

Key Features

  • Compliance with IEC 60601-1:2005 Ed 3.0, JIS T 0601-1:2012 for medical-use electrical devices and essential to electrical safety (*1)
  • Compliance with Electrical Appliances and Materials Safety Act, JIS, IEC, and UL standards for general-use electrical devices
  • Uninterrupted polarity switching function dramatically reduces cycle time
  • Support for rated currents up to 20 A gives the instrument more than adequate capability for testing products designed to comply with new standards
  • Communications functionality and external I/O support allow automatic testing on production lines

Basic specifications

Accuracy guaranteed: 1 year

Measurement methods Measurement of voltage drop across body simulated resistance points, Calculation and display of current values, True rms measurement, Measurement unit floats relative to instrument ground.
Measurement modes Leak current measurement, voltage measurement, safety conductor current measurement
Standards compliance
(NW: Body simulated resistance)
[NW-A] • Electrical Appliances and Materials Safety Act
[NW-B1] • Medical electrical equipment: IEC 60601-1:1988+ A1:1993+ A2:1995, JIS T 0601-1:1999
[NW-B2] • Medical electrical equipment: IEC 60601-1:2005+ A1:2012, JIS T 0601-1:2012 and complement 1:2014, IEC 62353
[NW-C]
• Measurement of touch current and protective conductor current: IEC 60990:2016
• Electrical equipment for measurement, control, and laboratory use: IEC 61010-1:2010+ A1:2016
• Information technology equipment: IEC60950-1:2005+ A1:2009+ A2:2013
• Audio, video and similar electronic apparatus: IEC 60065:2014
• Personnel Protection Systems for EV: UL 2231-1:2012 (Amended 2016), UL-2231-2:2012 (Amended 2016)
[NW-D] • For UL: UL 1492:1996 (Amended 2013)
[NW-G] • Electrical equipment for measurement, control, and laboratory use; current measurement circuits in damp conditions: IEC 61010-1:2010+ A1:2016
Leak current measurement Ground leak current, 3 types of contact current, 7 types of patient leak current, patient measurement current, 4 types of total patient leak current, free current measurement, 3 types of enclosure leak current
Measurement current DC, AC (true rms, 0.1 Hz to 1 MHz), AC+DC (true rms, 0.1 Hz to 1 MHz), AC peak (15 Hz to 1 MHz)
Measurement ranges DC / AC / AC+DC mode: 50.00 mA/ 5.000 mA/ 500.0 μA/ 50.00 μA
AC peak mode: 75.0 mA/ 10.00 mA/ 1.000 mA/ 500.0 μA
Measurement accuracy
(current measurement)
DC measurement: ±2.0% rdg. ±6 dgt. (typ.)
AC / AC+DC measurement: ±2.0% rdg. ±6 dgt. (15 Hz to 100 kHz, typ.)
AC peak measurement: ±2.0% rdg. ±6 dgt. (15 Hz to 10 kHz, typ.)
Interfaces External I/O, medical device relay output, USB 1.1 (communications), RS-232C
Functionality 110% voltage application, automatic test, data storage for 100 target devices, clock, data backup, printed output (optional), etc.
Power supply 100/120/220/240 V AC (specify at time of order), 50/60 Hz, 30 VA rated power
Target device power supply input 100 to 250 V AC, 50/60 Hz Rated current input from terminal block: 20 A
Target device power supply output Output from terminal block: 20 A Output from outlet: 15 A
Dimensions and mass 320 mm (12.60 in)W × 110 mm (4.33 in)H × 253 mm (9.96 in)D, 4.5 kg (158.7 oz)
Included accessories Test lead L2200 (for ST5540, Red ×2, Black ×1) ×1 set, Enclosure probe 9195 ×1, Power cord ×3, Spare fuse for measurement line ×1, Instruction manual ×1, CD-ROM ×1

ST5540, ST5541 List of functions (Network)

Item ST5540 ST5541
Network A
(Electrical Appliances and Materials Safety Act)
Network B (Medical-use electrical devices) -
Network C (IEC 60990)
Network D (UL)
Network E (General-purpose 1)
Network F (General-purpose 2)
Network G (IEC 61010-1)

List of functions (Major functions)

Item ST5540 ST5541
Power on polarity switching function
Rated current 20 A
Function for checking for blown fuses
Frequency band switching -
110% voltage output terminal (T3 terminal) -
S10, S12, S13, E terminal -

List of functions (Testing leakage current mode)

Item ST5540 ST5541
Earth leakage current
Touch current
Patient auxiliary current -
Patient leakage current -
Total patient leakage current -
Free current
Enclosure - Earth leakage current
Enclosure - Enclosure leakage current
Enclosure - Line leakage current
Patient leakage current I -
Patient leakage current II -
Patient leakage current III -
Included accessories Test lead L2200 (for ST5540, Red ×2, Black ×1) ×1 set, Enclosure probe 9195 ×1, Power cord ×3, Spare fuse for measurement line ×1, Instruction manual ×1, CD-ROM ×1
Included accessories Test lead L2200 (for ST5540, Red ×2, Black ×1) ×1 set, Enclosure probe 9195 ×1, Power cord ×3, Spare fuse for measurement line ×1, Instruction manual ×1, CD-ROM ×1

To ensure the safe use of electrical products, electrical safety tests such as insulation resistance, withstand voltage, ground resistance, and leakage current must always be conducted. Hioki Leakage Current Testers comply with laws and standards regarding medical electrical equipment as well as non-medical electrical equipment, and can be used to measure leakage current in all types of electrical products from computers to medical equipment. The ST5540 consists of measurement networks that simulate the human body and a high-frequency volt meter, and is also capable of switching power supply polarities and making measurements under simulated single-fault conditions in equipment under test. HIOKI LEAK CURRENT HiTESTER ST5540 Price in Bangladesh is competitive for precision insulation and leakage current testing. Ideal for industrial use, it ensures high accuracy and safety. Explore top features and reliable performance tailored for the Bangladeshi market. Get the best value with trusted quality and efficiency for your applications.

Key Features

  • Compliance with IEC 60601-1:2005 Ed 3.0, JIS T 0601-1:2012 for medical-use electrical devices and essential to electrical safety (*1)
  • Compliance with Electrical Appliances and Materials Safety Act, JIS, IEC, and UL standards for general-use electrical devices
  • Uninterrupted polarity switching function dramatically reduces cycle time
  • Support for rated currents up to 20 A gives the instrument more than adequate capability for testing products designed to comply with new standards
  • Communications functionality and external I/O support allow automatic testing on production lines

Basic specifications

Accuracy guaranteed: 1 year

Measurement methods Measurement of voltage drop across body simulated resistance points, Calculation and display of current values, True rms measurement, Measurement unit floats relative to instrument ground.
Measurement modes Leak current measurement, voltage measurement, safety conductor current measurement
Standards compliance
(NW: Body simulated resistance)
[NW-A] • Electrical Appliances and Materials Safety Act
[NW-B1] • Medical electrical equipment: IEC 60601-1:1988+ A1:1993+ A2:1995, JIS T 0601-1:1999
[NW-B2] • Medical electrical equipment: IEC 60601-1:2005+ A1:2012, JIS T 0601-1:2012 and complement 1:2014, IEC 62353
[NW-C]
• Measurement of touch current and protective conductor current: IEC 60990:2016
• Electrical equipment for measurement, control, and laboratory use: IEC 61010-1:2010+ A1:2016
• Information technology equipment: IEC60950-1:2005+ A1:2009+ A2:2013
• Audio, video and similar electronic apparatus: IEC 60065:2014
• Personnel Protection Systems for EV: UL 2231-1:2012 (Amended 2016), UL-2231-2:2012 (Amended 2016)
[NW-D] • For UL: UL 1492:1996 (Amended 2013)
[NW-G] • Electrical equipment for measurement, control, and laboratory use; current measurement circuits in damp conditions: IEC 61010-1:2010+ A1:2016
Leak current measurement Ground leak current, 3 types of contact current, 7 types of patient leak current, patient measurement current, 4 types of total patient leak current, free current measurement, 3 types of enclosure leak current
Measurement current DC, AC (true rms, 0.1 Hz to 1 MHz), AC+DC (true rms, 0.1 Hz to 1 MHz), AC peak (15 Hz to 1 MHz)
Measurement ranges DC / AC / AC+DC mode: 50.00 mA/ 5.000 mA/ 500.0 μA/ 50.00 μA
AC peak mode: 75.0 mA/ 10.00 mA/ 1.000 mA/ 500.0 μA
Measurement accuracy
(current measurement)
DC measurement: ±2.0% rdg. ±6 dgt. (typ.)
AC / AC+DC measurement: ±2.0% rdg. ±6 dgt. (15 Hz to 100 kHz, typ.)
AC peak measurement: ±2.0% rdg. ±6 dgt. (15 Hz to 10 kHz, typ.)
Interfaces External I/O, medical device relay output, USB 1.1 (communications), RS-232C
Functionality 110% voltage application, automatic test, data storage for 100 target devices, clock, data backup, printed output (optional), etc.
Power supply 100/120/220/240 V AC (specify at time of order), 50/60 Hz, 30 VA rated power
Target device power supply input 100 to 250 V AC, 50/60 Hz Rated current input from terminal block: 20 A
Target device power supply output Output from terminal block: 20 A Output from outlet: 15 A
Dimensions and mass 320 mm (12.60 in)W × 110 mm (4.33 in)H × 253 mm (9.96 in)D, 4.5 kg (158.7 oz)
Included accessories Test lead L2200 (for ST5540, Red ×2, Black ×1) ×1 set, Enclosure probe 9195 ×1, Power cord ×3, Spare fuse for measurement line ×1, Instruction manual ×1, CD-ROM ×1

ST5540, ST5541 List of functions (Network)

Item ST5540 ST5541
Network A
(Electrical Appliances and Materials Safety Act)
Network B (Medical-use electrical devices) -
Network C (IEC 60990)
Network D (UL)
Network E (General-purpose 1)
Network F (General-purpose 2)
Network G (IEC 61010-1)

List of functions (Major functions)

Item ST5540 ST5541
Power on polarity switching function
Rated current 20 A
Function for checking for blown fuses
Frequency band switching -
110% voltage output terminal (T3 terminal) -
S10, S12, S13, E terminal -

List of functions (Testing leakage current mode)

Item ST5540 ST5541
Earth leakage current
Touch current
Patient auxiliary current -
Patient leakage current -
Total patient leakage current -
Free current
Enclosure - Earth leakage current
Enclosure - Enclosure leakage current
Enclosure - Line leakage current
Patient leakage current I -
Patient leakage current II -
Patient leakage current III -

Included accessories Test lead L2200 (for ST5540, Red ×2, Black ×1) ×1 set, Enclosure probe 9195 ×1, Power cord ×3, Spare fuse for measurement line ×1, Instruction manual ×1, CD-ROM ×1

Included accessories Test lead L2200 (for ST5540, Red ×2, Black ×1) ×1 set, Enclosure probe 9195 ×1, Power cord ×3, Spare fuse for measurement line ×1, Instruction manual ×1, CD-ROM ×1
seller

Sellers

Connect your real buyer on Tools Bangaldesh where visit the maximum customer for Tools In Bangladesh.

e-marketplace

e-Marketplace

Tools Bangladesh is a Trustworthy e-Marketplace in Bangladesh. Connect buyer and seller on the trade fair.

growth

Buyers

Get the reliable and authentic Tools at the best price in Bangladesh with on time delivery entire Bangladesh.