The Hioki SS7012 is a compact DC signal source and calibrator that is used to perform maintenance or loop tests of instrumentation systems and can be operated from a computer via a USB cable. It can also be used to calibrate thermometers and measure DC voltage and DC current. HIOKI DC SIGNAL SOURCE SS7012 Price in Bangladesh offers excellent value for precise and reliable DC signal generation. Ideal for testing and calibration, this device ensures accurate performance in various applications. Its competitive price and trusted brand make it a top choice for professionals seeking quality signal sources.
Key Features
- Improve stability and reduce calibration costs compared with the previous HIOKI model
- For instrumentation systems (4 - 20 mA) and loop testing
- Check temperature control equipment and electric distribution
- 8 types of thermocouples to test thermoelectric power generation
- Ideal for electrical device evaluating and routine maintenance of production equipment such as calibrators
- Use the max. 25 mA DC sink as an electric load
Basic specifications [Generation functions]
Accuracy guaranteed: 1 year
| Circuit method |
Bipolar sink and source |
| Constant voltage |
2.5 V: 0 to ±2.5000 V (±0.03 % of setting ±300 μV, 100 μV resolution)
25 V: 0 to ±25.000 V (±0.03 % of setting ±3 mV, 1 mV resolution) |
| Constant current |
25 mA: 0 to ±25.000 mA (±0.03 % of setting ±3 μA, 1 μA resolution) |
| Thermoelectric power generation |
K: at TC: 0 ˚C, -174.0 to 1372.0 ˚C (±0.05 % of setting ±0.5 ˚C, 0.1 ˚C resolution), Other types: E, J, T, R, S, B, N selectable |
| Thermoelectric power generation |
K: at TC: RJ, -174.0 to 1372.0 ˚C (±0.05 % of setting ±1.0 ˚C, 0.1 ˚C resolution), Other types: E, J, T, R, S, B, N selectable |
| Standard resistance (Rs) |
100 Ω (±0.2 Ω) |
| Automatic generation |
Number of memory steps: 20, Interval time: 1 to 99 sec (at CV, CC, TC mode) |
[Measurement functions]
| Voltage |
2.5 V: 0 to ±2.8000 V (±0.03 % rdg. ±300 μV, 100 μV resolution, 1 MΩ input resistance)
25 V: 0 to ±28.000 V (±0.03 % rdg. ±3 mV, 1 mV resolution, 1 MΩ input resistance) |
| Current |
25 mA: 0 to ±28.000 mA (±0.03 % rdg. ±3 μA, 1 μA resolution, smaller than 30 Ω input resistance) |
| Temperature |
-25.0 to 80.0 ˚C (±0.5 ˚C at 23 ±5 ˚C, 0.1 ˚C resolution, use with the RJ sensor 9184) |
| Sampling rate |
Approx. 1.67 times/sec |
[General functions]
| Additional functions |
Zero adjustment, Overflow display, USB communication, Monitor |
| Power supply |
AC adapter 9445-02/-03 (100 to 240 V AC 50/60 Hz, 9 VA),
Ni-MH battery HR6 × 4, 6 VA, (fully charged 2500 mAh Ni-MH batteries: 170 minutes continuous use), or LR6 (AA) alkaline battery × 4, 6 VA |
| Dimensions and mass |
104 mm (4.09 in) W × 180 mm (7.09 in) H × 58 mm (2.28 in) D, 660 g (23.3 oz) (including LR6 × 4 batteries) |
| Included accessories |
Input cord 9168 ×1, Test lead L9170-10 ×1, Fuse ×1, LR6 (AA) alkaline battery ×4, Instruction manual ×1 |
| Included accessories |
Input cord 9168 ×1, Test lead L9170-10 ×1, Fuse ×1, LR6 (AA) alkaline battery ×4, Instruction manual ×1 |
| Included accessories |
Input cord 9168 ×1, Test lead L9170-10 ×1, Fuse ×1, LR6 (AA) alkaline battery ×4, Instruction manual ×1 |