Hioki flat bandwidth current probes are best-in-class current sensors for use with Memory HiCorders and high performance oscilloscopes. The CT6711 is a 3-range, high-sensitivity current probe that offers a wide DC to 120 MHz bandwidth and 200 μA to 30 A input, letting your capture signals across an expansive current range, from steady state to transient currents, when developing switching power supplies, inverters, and motor controllers. HIOKI CURRENT PROBE CT6711 Price in Bangladesh is ideal for measuring high-speed current waveforms with precision. This advanced probe ensures reliability and accuracy in electrical testing applications. Designed for professionals, it offers superior performance in compact form. Choose quality and check the HIOKI CURRENT PROBE CT6711 Price in Bangladesh today.
Key Features
- 3 ranges in a single probe - 30 A, 5 A, 0.5 A. Observe a wide current range from micro currents to 30 A.
- Wide band: DC to 120 MHz (-3 dB)
- High S/N ratio and 10 times output rate: Observe waveforms at 100 μA/div at oscilloscope maximum voltage sensitivity setting of 1 mV/div
- Directly connect to an oscilloscope’s BNC input terminal (*1)
Basic specifications
Accuracy guaranteed: 1 year
| Frequency bandwidth |
DC to 120 MHz (-3 dB) |
| Rise time |
2.9 ns or shorter |
Delay time
(Typical) |
30 A range: 12 ns, 5 A range: 12 ns, 0.5 A range: 13 ns
(Delay time relative to rising waveform of input signal 1 ns) |
| Noise level |
75 μA rms max (at 0.5 A range, using 20 MHz band measuring instrument) |
| Max. rated current |
30 A range: 30 A rms, 5 A range: 5 A rms, 0.5 A range: 0.5 A rms
(DC, and sine wave, requires derating at frequency) |
| Max. allowable peak current |
30 A range: ±50 A peak (within the input limit time 2 s)
5 A range: ± 7.5 A peak,
0.5 A range: ± 0.75 A peak (< 10 MHz), ±0.3 A peak (≥ 10 MHz) |
| Amplitude accuracy |
30 A range: ±3.0% rdg. ±1 mV, (Typical) ±1.0% rdg. ±1 mV (≤ 10 Arms, DC, 45 to 66 Hz sine wave, within the maximum peak current of each range)
5 A range: ±3.0% rdg. ±1 mV, (Typical) ±1.0% rdg. ±1 mV (DC, 45 to 66 Hz sine wave, within the maximum peak current of each range)
0.5 A range: ±3.0% rdg. ±10 mV, (Typical) ±1.0% rdg. ±10 mV (DC, 45 to 66 Hz sine wave, within the maximum peak current of each range) |
| Output rate |
30 A range: 0.1 V/A, 5 A range: 1 V/A, 0.5 A range: 10 V/A, (The output of this probe is internally terminated) |
| Measurable conductors |
φ 5 mm (0.20 in), Insulated conductor |
| Power supply |
Supplied from Power Supply 3269, Probe Power Unit Z5021 |
| Cable length |
Sensor cable (between relay box and sensor): 1.5 m (4.92 ft)
Power cable: 1.0 m (3.28 ft) (Power plug: FFA.0S.304.CLAC37Y / LEMO inc.) |
| Dimensions and mass |
Sensor: 155 mm (6.10 in)W × 18 mm (0.71 in)H × 26 mm (1.02 in)D, Relay box section: 45 mm (1.77 in)W × 120 mm (4.72 in)H × 25 mm (0.98 in)D
Terminator section: 29 mm (1.14 in)W × 83 mm (3.27 in)H × 40 mm (1.57 in)D mm, 370 g (13.1 oz) |
| Included accessories |
Instruction manual ×1, Carrying case ×1 |
| Included accessories |
Instruction manual ×1, Carrying case ×1 |
| Included accessories |
Instruction manual ×1, Carrying case ×1 |