Fluke ProSim 2 and 3 Patient Simulator
Don’t need a comprehensive patient monitor tester? The 6-in-1 ProSim 3 and 4-in-1 ProSim 2 Vital Signs Simulators are clear choices for biomedical engineers and field service technicians that need a quality, feature-rich device with high portability. Choose one of these modern vital signs simulators for preventive maintenance, troubleshooting and repair. The ProSim 3 and 2 feature the perfect amount of features for testing in the field. We like to call it the Just Right feature set, and it includes: • ECG • Pacemaker • Arrhythmia and performance testing • Respiration • Invasive blood pressure • Temperature • Cardiac output (ProSim 3 only) • Fetal/maternal (ProSim 3 only
• Portable, for evaluating the performance of patient monitors in the field
• 20 % lighter and 25 % smaller than preceding technology
• Just Right feature set includes: ECG, pacemaker, arrhythmia and performance testing, respiration, invasive blood pressure, temperature, cardiac output (ProSim 3 only), fetal/maternal (ProSim 3 only)
• 43 high-quality waveforms
• With four IBP channels, ProSim 3 tests even the highest acuity scenarios
• Stay-connected ECG posts for secure lead connections
• Improved user interface and online Advantage Training demos
• Upgraded DIN connectors ensure consistency with the ProSim family; minimize cable compatibility issues
• Field upgradeable, and easily paired with other devices for comprehensive testing
• ProSim 3 and 2 are 510(k) cleared products
General Specifications
-
Operating Temperature: 10°C to 40°C (50°F to 104°F)
-
Storage Temperature: -25°C to 50°C (-13°F to 122°F)
-
Humidity: 10% to 80% non-condensing
-
Altitude: Up to 2,000 meters (6,562 ft)
-
Dimensions: 14.0 cm x 20.6 cm x 4.5 cm (5.5 in x 8.2 in x 1.8 in)
-
Display: LCD greyscale
-
Communication: USB device upstream port
-
Power: Two 9 V alkaline batteries
-
Battery Life: 8 hours continuous operation
-
Weight: 0.47 kg (1 lb, 4 oz)
-
Safety Standards: IEC 61010-1, Pollution degree 2
-
Certifications: CE, CSA, C-TICK N 10140, RoHS
-
Electromagnetic Compatibility (EMC): IEC 61326-1; 2006
Detailed Specifications
ECG Waveform
-
Amplitude: 0.05 mV to 0.45 mV (0.05 mV steps); 0.5 mV to 5.5 mV (0.5 mV steps)
-
Amplitude Accuracy: ±2% of setting for Lead II; ±5% for other leads
-
Rate: 30 to 300 BPM in 20 BPM increments
-
Rate Accuracy: ±1% of setting
-
QRS Duration: Adult (80 ms) or pediatric (40 ms)
-
ST-Segment Elevation: -0.8 mV to +0.8 mV (0.1 mV steps)
-
Power-On Default: 80 BPM, 1.0 mV, adult QRS, 0 mV ST-segment elevation, 0.16-second P-R interval
Pacemaker Waveform
-
Pacer Pulse Amplitude: 0 (off), 1, 2, 5, 10 mV ±10% for Lead II
-
Accuracy: ±(5% setting + 0.2 mV) for Lead II
-
Pulse Width: 0.1, 0.5, 1.0, 1.5, 2.0 ms ±5%
Arrhythmia
-
Baseline: Normal Sinus Rhythm (NSR) at 80 BPM
-
PVC Focus: Left focus, standard timing
-
Supraventricular Arrhythmias: Atrial fibrillation (coarse or fine), atrial flutter, sinus arrhythmia, missed beat (one time), atrial tachycardia, paroxysmal atrial tachycardia, nodal rhythm, supraventricular tachycardia
-
Premature Arrhythmias: Premature atrial contraction (PAC), premature nodal contraction (PNC), PVC1 left ventricular, PVC1 left ventricular early, PVC1 left ventricular R on T, PVC2 right ventricular, PVC2 right ventricular early, PVC2 right ventricular R on T, multifocal PVCs
-
Ventricular Arrhythmias: PVCs 6, 12, or 24 per minute; frequent multifocal PVCs; bigeminy; trigeminy; multiple PVCs (one-time run of 2, 5, or 11 PVCs); ventricular tachycardia; ventricular fibrillation (coarse or fine); asystole
-
Conduction Defects: First-, second-, or third-degree AV block; right- or left-bundle branch block