Fluke ESA710 Electrical Safety Analyzer
Redefining automated testing in a compact format, the ESA710 Electrical Safety Analyzer is an essential tool for healthcare technology management professionals looking for portability without compromising functionality. This high-functioning device enables users to perform from basic troubleshooting to comprehensive analysis, ensuring electrical safety testing on medical equipment complies with key electrical safety standards, such as ANSI/AAMI ES1 and NFPA 99, whether on-field or within facilities. Buy Fluke ESA712 Electrical Safety Analyzer at the best price in Bangladesh. Reliable, accurate, and perfect for all your electrical safety testing needs.
• 5” Touchscreen with intuitive user interface
• Small, portable, ergonomic design with an integrated handle
• OneQA-enabled workflow automation eliminates the need for a laptop
• Wireless capability allows you to test procedures in the field and automatically sync results
• Ten applied parts posts with innovative LED lighting guidance (ECG & Banana)
• Insulation resistance from 50V, 100V, 250V and 500V
• On-board automation with automated test sequences for compliance with key electrical safety standards (ANSI/ AAMI ES1, NFPA 99)
• ECG waveform tests and dual-lead measurements combine the functionality of a simulator and safety analyzer in a single test tool
• Ability to upload pictures, instructions, and text to procedures
• Easy data entry through barcode scanner, external keyboard, or on-screen keyboard
• 20 A at 120 V equipment current capability
• Replaceable mains fuses keep your unit in the field and out of the repair shop
• Rigorously tested for rugged field applications with CE and CSA in addition to Fluke-quality ruggedness for long-term reliability
• Enjoy precision, compliance, and total peace of mind through our comprehensive service and coverage program, eliminating unexpected costs with our Protect+ Program
• Global support network delivering prompt service and peace of mind to Fluke Biomedical customers worldwide
| |
General Specifications
| Parameter |
Specification |
| Safety Standard Compliance |
IEC 61010-1: Overvoltage category II, pollution degree 2
IEC 61010-2-034: Measurement CAT II 300 V |
| Dimensions (W x D x H) |
214 x 207 x 92 mm (8.4 x 8.1 x 3.6 in) |
| Weight |
1.3 kg (2.7 lbs) |
| Ingress Protection |
IP40 (excluding equipment outlet) |
| Operating Temperature |
0 to +35 °C (50 to 95 °F) |
| Operating Humidity |
10–90%, non-condensing |
| Storage Temperature |
-20 to +60 °C (-4 to +140 °F) |
| Storage Humidity |
5–95%, non-condensing |
| Altitude |
≤5000 m @ 100–127 V, ≤2000 m @ 200–240 V |
| Battery Life |
Up to 2 hours |
| Connectivity |
1× USB-C (PC), 2× USB-A (peripherals) |
| Display |
5-inch touchscreen |
| Data Storage |
>10,000 measurements |
| Power |
Multiple ranges:
• 90–132 V AC, 20 A MAX, 47–63 Hz
• 90–132 V & 180–264 V AC, 15 A or 16 A MAX |
| CSA Approval |
Yes (US version and NEMA 6-15 version) |
| EMC |
See user manual |
Measurement Specifications
| Category |
Parameter |
Specification |
| Mains Voltage |
Range |
0–264 V AC RMS |
| |
Accuracy |
±(2% + 0.2 V) |
| Point to Point Voltage |
Range |
0–300 V AC RMS |
| |
Accuracy |
±(2% + 0.2 V) |
| Earth & Point-to-Point Resistance |
Range |
0–20 Ω |
| |
Accuracy |
±(1% + 0.01 Ω) at ≤2 Ω, ±(1% + 0.1 Ω) at >2 Ω |
| |
Test Current |
Square wave ±200 mA |
| |
Open Circuit Voltage |
Max ±24 V |
Equipment Current
| Parameter |
Specification |
| Range |
0–20 A AC RMS |
| Accuracy |
±(5% + 0.05 A) |
| Duty Cycle |
0–10 A: continuous
0–15 A: 5 min on / 5 min off
15–20 A: 3 min on / 7 min off |
Leakage Current
| Parameter |
Specification |
| Modes |
AC + DC (True RMS), AC only, DC only |
| Crest Factor |
≤2 |
| Range |
0 µA – 20 mA |
| Accuracy |
- DC & 20 Hz–0.5 kHz: ±(1% + 1 µA)
- 0.5–50 kHz: ±(2.5% + 1 µA)
- 50 kHz–1 MHz: ±(5% + 1 µA)
|
| Main Test Voltage |
Mains ±5% |
| Current Limit |
1 mA ± 25% at 115 V |
| Uncertainty |
±2 µA at 120 V, ±4 µA at 230 V |
Insulation Resistance
| Parameter |
Specification |
| Test Voltages |
50 V, 100 V, 250 V, 500 V DC |
| Range (250/500 V) |
0.1–100 MΩ |
| Accuracy |
±(2% + 0.2 MΩ) at ≤10 MΩ
±(7.5% + 0.2 MΩ) at >10 MΩ |
| Range (50/100 V) |
0.1–20 MΩ |
| Accuracy |
±(10% + 0.2 MΩ) |
| Voltage Accuracy |
+20%/-0% (250/500 V), +30%/-0% (50/100 V) |
| Short-Circuit Current |
2 mA ± 0.25 mA |
| Max Load Capacitance |
2 µF |
ECG Simulation
| Parameter |
Specification |
| Frequency Accuracy |
±2% |
| Amplitude Accuracy |
±5% (for 2 Hz square wave) |
| Waveforms |
ECG complex (30–240 BPM), Square (0.125/2 Hz), Triangle (2 Hz), Pulse (30/60 BPM), Ventricular fibrillation |
Respiration Simulation
| Parameter |
Specification |
| Rate |
Apnea (0 BrPM) and 10–100 BrPM in 10 BrPM steps |
| Waveform |
Normal |
| Inspiration: Expiration Ratio |
1:1 |
| Impedance Baseline |
1000 Ω ± 5% |
| <
strong>Impedance Variation |
1 ± 0.15 Ω |
| Respiration Lead |
LL or LA (user selectable) |
|
| Standard Accessories |
|
| Getting started guide |
5517611 |
| OneQA quick guide |
5517610 |
| USB C to USB A cable |
5579600 |
| USA accessory kit |
3111008 (includes test lead set, TP1 test probe set, AC285 alligator clip set) |
| Probe test banana jack 4MM tip black w/cap |
2427138 |
| Probe test banana jack 4MM tip red w/cap |
2099044 |
| Null post adapter |
3326842 |
| Carrying case |
2248650 |
| Power cord |
2238644 |
| Optional Accessories |
|
| Barcode scanner |
6078258 |
| USB Wi-Fi dongle |
6078184 |