Fluke ESA612 Electrical Safety Analyzer
The ESA612 Electrical Safety Analyzer represents the next generation in test equipment for biomedical and clinical engineering professionals who perform field service on medical equipment throughout their facilities, in clinics, and anywhere onsite service is required. Portable, lightweight, and designed for operation in tight spaces, the ESA612 offers the functionality of a simulator, multimeter, and electrical safety tester in a single test tool. With the selection of two test loads, this versatile product can be used worldwide for compliance testing to preventative maintenance of electrical safety standards of choice. The versatility of the multifaceted ESA612 tester is further expanded with optional automation software, which speeds and simplifies testing and provides high-end analyzer productivity at a software level investment. Optional Ansur-automated ESA612 software standardizes test procedures, compares results to standard limits, and generates and stores reports for total digital data management. Find the best Fluke ESA612 Electrical Safety Analyzer price in Bangladesh. Accurate, portable, and ideal for safety testing.
- Portable, ergonomic, lightweight, and easy to use
- Large, easy-to-read display with adjustable contrast
- Human-factors-designed user interface
- Tilt stand design for stand-up testing in field environments
- Five applied parts jacks and easy ECG snap connection with optional expander box
- ECG waveform tests and dual-lead measurements combine the functionality of a simulator, multimeter and safety analyzer in a single test tool
- Replaceable mains fuses keep your unit in the field and out of the repair shop
- Internal memory for 100 test records
- 20 A at 120 V current capability
- USB connection for use with Ansur and Data Viewer software (for memory download to PC)
- Optional Ansur automation software standardizes test procedures, compares results to standards limits, generates/stores reports and provides total digital data management
- Rigorously tested for rugged field applications, with CE and CSA in addition to the Fluke-quality-design stamp of approval
|
| Voltage |
| Range (mains voltage) |
180 to 264 V ac rms |
| Range (accessible voltage) |
0 V ac to 300 V ac rms |
| Accuracy |
± (2 % of reading + 0.2 V) |
| Voltage tests |
Mains, accessible, and point-to-point |
| Earth resistance |
| Mode |
Two terminal |
| Test current |
> 200 mA ac |
| Range |
0 W to 2 W |
| Accuracy |
± (2 % of reading + 0.015 W) |
| Resistance tests |
Earth resistance and point-to-point |
| Equipment current |
| Mode |
AC rms |
| Range |
0 A to 20 A |
| Accuracy |
± 5 % of reading + (2 counts or 0.2 A, whichever is greater) |
| Duty cycle |
15 A to 20 A, 5 min on/5 min off
10 A to 15 A, 7 min on/3 min off
0 A to 10 A continuous |
| Leakage current |
Modes
Modes are available in all leakage tests with the exception
of MAP leakages that are available only in true-rms |
AC + DC (true-rms)
AC only
DC only |
Patient load selection
(input impedance) |
AAMI ES1-1993
IEC 60601 |
| Crest factor |
= 3 |
| Ranges |
0 µA to 199.9 µA
200 µA to 1999 µA
2 mA to 10 mA |
Frequency response/accuracy
DC to 1 kHz
1 kHz to 100 kHz
100 kHz to 1 MHz |
± (1 % of reading + (1 µA or 1 LSD, whichever is greater))
± (2 % of reading + (1 µA or 1 LSD, whichever is greater))
± (5 % of reading + (1 µA or 1 LSD, whichever is greater)) |
| Leakage tests |
Ground wire (earth)
Chassis (enclosure)
Lead to ground (patient)
Lead to lead (patient auxiliary)
Lead isolation (mains on applied part)
Direct equipment
Direct applied part
Alternative equipment
Alternative applied part
Point to point |
| Mains on applied part test voltage |
100 % of mains |
| Differential leakage |
| Ranges |
10 µA to 199 µA
200 µA to 2000 µA
2 mA to 20 mA |
| Accuracy |
± 10 % of reading + (2 counts or 20 µA, whichever is greater) |
| Insulation resistance |
| Ranges |
0.5 MW to 20 MW
20 MW to 100 MW |
| Accuracy |
± (2 % of reading + 0.2 MW)
± (7.5 % of reading + 0.2 MW) |
| Source test voltage |
500 V dc
250 V dc |
| Insulation resistance tests |
Mains-PE, AP-PE, Mains- PE, Mains-NE (non-earthed accessible
conductive part) and AP- NE (non-earthed accessible conductive
part) |
| ECG performance waveforms |
| Accuracy |
± 2 %
± 5 % for amplitude of 2 Hz square wave only, fixed at 1 mV
Lead II configuration |
Waveforms
ECG complex (BPM)
Ventricular fibrillation
Square wave (50 % duty cycle) (Hz)
Sine wave (Hz)
Triangle wave (Hz)
Pulse (63 ms pulse width) |
30, 60, 120, 180, and 240
0.125 and 2
10, 40, 50, 60, and 100
2
30 BPM and 60 BPM |
| Power ratings |
| Mains voltage outlet |
230 V ac |
| Mains voltage inlet power range |
180 to 264 V ac rms |
| Maximum current |
16 A |
| Hz |
50 or 60 |
| Environmental specifications |
| Operating temperature |
10 °C to 40 °C (50 °F to 104 °F) |
| Storage temperature |
-20 °C to 60 °C (-4 °F to 140 °F) |
| Operating humidity |
10 % to 90 % non-condensing |
| Altitude |
230 V ac mains supply voltage up to 2,000 m |